beautypg.com

Cooper Instruments & Systems Model 7i Professional Force/Torque Indicator User Manual

Page 31

background image

Model 7i Digital Force / Torque Indicator

User’s Guide


30

Memory & Statistics

MEM

Transmit all stored readings

STA

Transmit

statistics


Set

Points

SPHD

Disable high set point

SPLD

Disable low set point

SPHn

High set point. n=value (+ for C/CW, - for T/CCW)

SPLn

Low set point. n=value (+ for C/CW, - for T/CCW)
Note: High set point value must be greater than low set point value if both values
are set to the same polarity.

USB/RS-232

Communication

FULL

USB/RS-232 transmission with units

NUM

USB/RS-232 transmission without units (only numeric values)

AOUTn

Auto-transmit n times per second n=1, 2, 5, 10, 25, 50, 125, 250. 0=disabled
Note: n = 1 = yields 50 times per second. This is provided for backward
compatibility with the legacy indicator model BGI.

IPOLn

Invert polarity of output. n=1=invert polarity. n=0=normal (default)

Note: Normal polarity is positive for compression and negative for tension.

OPOLn

Omit polarity of output. n=1=omit polarity. n=0=include polarity (default)

Note: The “+” sign is always omitted. A “–“ sign is sent when polarity is enabled.

Mitutoyo Communication

MIT

Enable Mitutoyo output

MITD

Disable Mitutoyo output

POL

Mitutoyo output with polarity (+ for compression, - for tension)

NPOL

Mitutoyo output without polarity (absolute value)

PM

Print/send data to a Mitutoyo-compatible device

Averaging

A

Enable Average mode

AD

Disable Average mode

AM

Select Average mode (if enabled) for primary reading

TRFn

Trigger load. n=value (+ for C/CW, - for T/CCW)

DELn

Initial delay. n=0.1-300.0 seconds

ATn

Average time. n=0.1-300.0 seconds


External Trigger

ETH

Enable high level-triggered External trigger mode

ETL

Enable low level-triggered External trigger mode

ETHL

Enable reading captured on a high to low transition

ETLH

Enable reading captured on a low to high transition

ETD

Disable External trigger mode

Input / Output Bits

Sn

Set output bit (open drain, pull to ground). n=0, 1, 2

Cn

Clear output bit. n=0, 1, 2

Rn

Read current status of output bit or level of input pin. n=0, 1, 2 ,3

Personality

RN

Read product name

RM

Read

model

number

RV

Read firmware version number

RS

Read serial number