Section 6
Bio-Plex 200 System Specifications
General Technical Specifications
Environmental conditions
Operating temperature
15–30°C (59–86°F)
Operating humidity
20–80%, noncondensing
Operating altitude
Designed to operate at 2,400 m (7,874 ft) above
mean sea level or below
Compensatory range
± 2°C
UL installation category
UL Installation Category II, as defined in Annex J of
UL 3101-1
Pollution degree
Pollution Degree 2, as defined in Section 3.7.3.2 of
UL 3101-1
Array Reader Specifications
Input voltage range
100–240 V, ~1.5 A, 47–63 Hz
Physical dimensions (W x D x H)
43 x 51 x 23 cm
Weight
23 kg (60 lb)
Lasers
Reporter laser
532 nm, 10 mW
Classification laser
635 nm, 8.5 mW, diode
Fluidics
Sheath flow rate
90 µl/sec
Cuvette
200 µm square flow channel
Sample injection rate
60 µl/min
Electronics
Reporter channel
Photomultiplier tube, A/D resolution 14 bits
detection
Classification and
Avalanche photodiodes with temperature
doublet discriminator
compensation, A/D resolution 12 bits
channel detection
Communications interface
RS232 and USB
Signal processing
Measurement resolution
15 bits effective
Processor modes
Linear, with logarithmic or linear display option
Dynamic range
70 dB
36