Bio-Rad Bio-Plex® 200 Systems User Manual
Page 36
Bio-Plex Manager
Incorrect bead regions
Compare bead regions in the assay
has detected a problem with
selected in the protocol
with those selected in the protocol.
bead selection.
Incorrect regions selected
Verify regions chosen during assay
when preparing the assay
preparation.
Too few beads in the assay
Verify that the correct number of
beads in one or more regions
were used.
Bio-Plex Manager has
Most likely:
detected a problem with
assignment of beads into
Calibration performed before
Perform 30 min Warm Up and
regions.
the array reader was warmed
recalibrate.
up
Improper calibration
Check that the target values of the
CAL beads match values entered in
the software, then recalibrate.
Microbubbles present in
Perform Remove Bubbles.
cuvette.
Less likely:
Calibration beads are
Recalibrate with new Cal1 beads.
photobleached (do not
expose to light for more
than 1 hr)
Array reader was calibrated
Clean MCV plate and recalibrate.
with a dirty MCV plate
Misalignment of optics
Perform Optical Validation.
Contact Technical Support if values
are not within range.
Bio-Plex Manager has
Clumped beads present
Vortex plate at 900 rpm for 1 min.
detected a problem with
aggregated beads.
Sheath reservoir is empty
Refill sheath reservoir. Perform
Start Up.
Waste reservoir is
Empty waste and reconnect.
overfilled
Problem with doublet
Use default DD gate setting. Run
discrimination (DD)
Classify Validation to check DD
efficiency.
Incompatible suspension
Check hardware manual for buffer
buffer used
compatibility.
Check Link in status bar of
Array reader or microplate
Turn on array reader and microplate
software.
platform not turned on
platform.
Software is not communicating Close and restart Bio-Plex Manager
with array reader
software.
Cables from computer to
Check cables for proper connections.
array reader or microplate
platform are loose/not
connected
31
www.bio-rad.com/bioplex/
Message/Problem
Causes Cause
Solution