Bio-Rad Bio-Plex® 200 Systems User Manual
Page 17
4.
The maximum distance between the computer and the microplate platform and array reader
should be 1.5 m (5 ft), the length of the communications cable supplied with the instruments.
5.
Do not place any items on top of the array reader. The cover is not designed to support
objects and thus the optics could be damaged.
6.
If installing the HTF, allow an area ~3 ft below the array reader for the 20 L sheath fluid
cube.
Note: The array reader contains sensitive optics that can be forced out of alignment through improper
handling and unnecessary moving. It is recommended that an authorized service representative
move your system. Following any system moves, it is necessary to validate the optical alignment
and report any changes. Refer to the Bio-Plex validation kit manual for validation of the optical
alignment.
3.3 Microplate Platform Setup
The microplate platform should be shipped with the following items:
•
Microplate platform
•
Power cord
•
Communication cable
•
Needle guide
•
2 sample needles (11.6 cm/4.6 in)
•
Shield
1.
Unpack all components and ensure that all accessories are supplied.
2.
Place the microplate platform on a clean, flat, and stable surface.
3.
Unscrew and remove the shipping pin (Figure 10).
Fig. 10. Remove shipping pin.
4.
Make sure that the power switch to the microplate platform is off, and connect the
communications cable to the microplate platform’s communication cable port. Attach a
power cord to the power connector of the microplate platform, but do not plug it into an
outlet until the array reader and microplate platform are aligned (Figure 11).
5.
Position the microplate platform in the location where you want it to be used.
6.
Connect the communication cable to communication port 1 on the rear of the computer.
3.4 Array Reader Setup
The array reader should be packaged with the following items:
•
Array reader
•
Power cord
12
www.bio-rad.com/bioplex/
Shipping pin