2 troubleshooting guide for vacuum manifold – Bio-Rad Bio-Plex® 200 Systems User Manual
Page 40
5.2 Troubleshooting Guide for Vacuum Manifold
Message/Problem
Causes
Solution
No Flow/no vacuum.
Lid on plate
Remove lid.
All wells not wet, or
Wet unused wells with Milli-Q
unused wells not covered
water. (You can reuse these wells
or sealed
later.) Or, tape the unused wells
with sealing tape. If you need to seal
partial rows or columns, seal the
unused rows or columns with tape
and leave the adjacent unused row
partially sealed. Then wet with buffer
since it is difficult to make seals at the
row edge.
Poor alignment of plates
Align plates.
with gasket
Vacuum trap filled
Empty trap.
Filter on pump clogged
Replace filter.
Pump not turned on
Turn on pump.
Manifold ON/OFF valve in
Turn to ON position.
OFF position
Manifold pressure gauge
Turn up to higher value.
turned to the lowest value
Damaged gasket
Replace gasket.
Gauge plug missing
Replace gauge plug in side of
manifold ring using hex key wrench.
Wells do not empty at the
Lid on plate
Remove lid.
same time/uneven flow.
Vacuum line turned off or
Clear line and repeat.
clogged
Debris in sample
Remove cellular debris prior to
adding to wells.
Bad plate-to-vacuum
To ensure a good plate-to-manifold
manifold seal
seal, put pressure on corners of
plate.
Too many beads
Check to ensure that proper bead
concentration was used.
Leakage during incubation.
High surfactant
Lower the concentration.
concentration
Failure to blot under-drain
Blot under-drain.
after filtration and before
incubation
Under-drain contacting
Place plate on smooth, flat surface
surface
so nothing touches the underdrain
spouts.
Absorbent material
Place on flat, nonabsorbent
contacting under-drain
material (such as lid).
Excessive agitation or
Mix using orbital table at a lower
vibration
volume (maximum 200 µl on shaker,
340 µl without a shaker) or use lower
speed.
35
www.bio-rad.com/bioplex/