Bio-Rad Bio-Plex® 200 Systems User Manual
Page 39
Blue power indicator light on
1. System not plugged in
Turn on system, power strip.
HTF is not on
2. Power strip not on
Remove power cable and reconne
HTF sheath reservoir does
Power not on
Press prime button on front of
not refill.
HTF.
Lines not connected
Turn power off then on again on
HTF.
Sheath cube more than
Disconnect air tubing
3 ft below HTF
connecting array reader to HTF.
HTF should prime and fill.
HTF audible alarm and red
Sheath cube is empty
Insert filtered end of sheath
fault indicator light on.
intake line into new container
and press Prime button on front
of HTF.
HTF not on level
Ensure that HTF system is on
surface
flat surface.
Filtered end of sheath
Ensure that filter is below level
tube not below level
of sheath in cube.
of sheath in cube
Sheath container is not
Place the sheath cube 3–4 ft
below level of HTF
below the level of the reader and
HTF system (Figure 20).
Sheath filter clogged
Replace sheath filter.
HTF may be overfilled
Disconnect the Sheath Out fitting
at the array reader. Discharge sheath
fluid into a waste container by
depressing the plunger at the end of
the Sheath Out tubing using a screw-
driver or other similar object. Drain
until HTF tank is about half full.
Call Technical Support if all actions
above do not resolve problem.
34
Message/Problem
Causes
Solution
www.bio-rad.com/bioplex/