beautypg.com

3 description of major components – Bio-Rad CHEF Mapper® XA System User Manual

Page 7

background image

Accessories Included

Variable speed oscillating pump 120 V, ground isolated. Flow rate 1 liter/min, typical

Casting stand

14 cm x 13 cm

Comb

15 well comb and comb holder

Tygon tubing

365 cm

Disposable Sample plug mold

50 slot

Yeast DNA standard

Saccharomyces cerevisiae YNN295

Chromosomal grade agarose

5 g

Pulsed field certified agarose

5 g

Leveling bubble

Manual

Cooling Module

Weight

14 kg

Construction

Aluminum

Dimensions

42 cm long x 23 cm wide x 24 cm high

Cooling capacity

75 watts of input power at 14 °C

Operating range

5–25 °C

Fuse

100/120 V: 6.3 amp, 250 Slo-blo
220/240 V: 3.1 amp, 250 Slo-blo

Total system weight

41.7 kg

1.3 Description of Major Components

Power Module

The power module contains the electronics for pulsed field electrophoresis, including a

350 V power supply, the switching functions, and drivers for the 24 electrodes. The front
panel contains a two line fluorescent display, buttons, switches, jacks, and a fuse as described
in Section 4. The fused power supply operates with a maximum of 9 V/cm, or 350 V. The
lowest gradient is 0.6 V/cm, or 20 V. The drivers provide clamped homogeneous electric
fields in the chamber and maintain them regardless of the pulse angle selected. This feature,
dynamic regulation, regulates the potentials so that the proper voltages are maintained
regardless of gel size or fluctuations in buffer conductivity or temperature.

Figure 1.1A shows the relative potentials of each electrode pair when the + 60° vector

(indicated by the arrow) is activated. Net field vector is from NW to SE. The highest poten-
tials are along the SE segment of the hexagon. The potentials gradually decline along the
adjacent segments. The NW segment, directly opposite the SE, has 0 potential, represented in
the diagram as negative terminals. When the - 60° angle is activated, the pattern of electric
charges is as shown in Figure 1.1B. Together, the two pulses result in a 120° included pulse
angle. Other angles will result in similar relative electrode potentials. One such example is field
inversion, shown in Figure 1.2A and 1.2B. In this case, the charges are primarily at N and S.
The pulse angles are 180° and 0°. Appropriately scaled voltages are along the sides of the
hexagon to achieve clamping.

3