Interactive computer algorithm for full optimization of electrophoresis conditions requires

PC 80386 or compatible, with Microsoft Windows

®

3.1. Buffer type, buffer concentration,

agarose concentration and type, and buffer temperature, can be varied as inputs.

Power Module

Dimensions

34.5 (depth) x 55.9 (width) x 30.5 (height) cm

Construction

Aluminum chassis

Weight

16 kg

Power supply

350 V maximum, to allow maximum gradient of 9 V/cm,
continuously adjustable; built in

Maximum current

0.5 amperes

Allowable voltage gradients

0, and 0.6–9 V/cm, in 0.1 V/cm increments

Battery back up

All parameters in memory, up to 3 hr of interruption

Delayed start

Up to 72 hr

Electrode potentials

Dynamically regulated (feedback adjustment) +/- 0.5% F.S.

Program storage

20 average protocols

Display

Fluorescent, 2 lines x 40 characters per line, alpha-numeric

Environmental

Operating

50 °F (10 °C) to 90 °F (32 °C) temperature 30-80%
humidity

Storage

32 °F (0 °C) to 140 °F (60 °C) temperature 10-90%
humidity

Switching Functions

Switching range

50 msec to 18 hr

Switch angle variable

0–360 degrees (all electronic switching) in 0.5° increments

Multistate vector switching

Up to 15 vectors per pulse cycle, each definable by angle,
voltage, and duration

Switch time ramps

Linear, concave, or convex using hyperbolic function

Secondary pulses

Defined by voltage, frequency, angle, and duration

Field inversion (FIGE)

Available with asymmetric forward, reverse voltages

Maximum program blocks

8, with automatic execution

Maximum run time

999 hours per block

Fuses

3 Amp Slo-Blo; two each for AC line input
0.5 Amp Fast Blo for high voltage output

Electrophoresis Cell

Dimensions

11.4 (h) x 44.2 (w) x 50.3 (d) cm, horizontal format

Construction

Cover

Vacuum formed polycarbonate

Base

Injection molded polycarbonate

Lid

Safety interlocked

Weight

10.2 kg

Electrodes

24, platinum (0.02 inch diameter)

Temperature monitoring

via precision temperature probe mounted in base of cell

2