Appendix d: product specifications and information, Product specifications – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual
Page 64

Gene Pulser MXcell™ System Manual
55
Appendix D: Product Specifications
and Information
The following tables provide product specifications and information about the Gene Pulser
MXcell™ electroporation system, including operating specifications and catalog numbers.
Product Specifications
Table 18 lists specifications for the entire system, including the power module and plate
chamber:
Table 19 lists the general specifications for the entire system:
Table 18. System specifications for Gene Pulser MXcell electroporation system
Waveform
Exponential decay or Square wave
Voltage
10–500 V in 2 V increments
Capacitance
25–2475 μF in 25 μF increments
Resistance
(Parallel)
50–1,500
Ω in 50 Ω increments plus infinity
Sample
Resistance
10
Ω minimum at 10–500 V; 125 Ω with Gene Pulser electroporation buffer
Square Wave
Timing
10–500 V: 0.05–9999.95 ms pulse length, 1–3 pulses
0.1–10 sec pulse interval
Table 19. General specifications for Gene Pulser MXcell electroporation system
Type
Specification
Input Voltage
100–120 VAC or 220–240 VAC, 50/60 Hz
Power
Max 240 W (During short charging periods)
Operating environment
Temperature 0–35
o
C, Humidity 0–95%, (non-condensing)
Regulatory
Safety EN 61010, EMC EN61326 Class A
Dimensions
Power Module 31 x 30 x 14 cm, weight 6.62 kg