beautypg.com

Mixed protocols, Opt 96 well/ exp,sqr, Opt 24 well/ exp,sqr – Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual

Page 47

background image

Gene Pulser MXcell™ System Manual | Pre-Set Protocols

38

Mixed Protocols

The following protocols include a mix of protocols with a variety of well sets, waveforms,
and plate sizes.

Opt 96 well/ Exp,Sqr

Table 13 shows the parameters that vary in well sets for a 96-well plate. The protocol varies
in a wide range of starting conditions for exponential and square waveforms.

Opt 24 well/ Exp,Sqr

These are parameters that vary in each well for a 24-well plate. Except for the larger well
size, the protocol is the same as “Opt 96 well/ Exp,Sqr” on page 38.

Table 13. Opt 96 well/ Exp, Sqr conditions that vary across the plate.

1

2

3

4

5

6

7

8

9

10

11

12

E
X
P

A

150 V
350
μF

200 V
350
μF

250 V
350
μF

300 V
350
μF

350 V
350
μF

450 V
350
μF

250 V
200
μF

250 V
250
μF

250 V
350
μF

250 V
500
μF

250 V
750
μF

250 V
1000
μF

B

C

D

S
Q
R

E

150 V
20 ms

200 V
20 ms

250 V
20 ms

300 V
20 ms

350 V
20 ms

450 V
20 ms

250 V
5 ms

250 V
10 ms

250 V
15 ms

250 V
20 ms

250 V
25 ms

250 V
30 ms

F

G

H