Gene Pulser MXcell™ System Manual | Pre-Set Protocols

34

Mini-Optimization Protocols

The following pre-set protocols use four or six well sets in the electroporation plate.

Opt mini 96 well/ Sqr.Exp

Table 7 shows the parameters that vary in each well set in a 96-well plate. The protocol
includes the following well sets:

Well set ABCD 1-3: Square wave, 200 V, 2,000 μF, 1,000

Ω, and 20 ms

Well set ABCD 4-6: Exponential wave, 250 V, 350–750 μF, and 1000

Ω

Table 7. Opt. mini 96/ Sqr.Exp conditions that vary across the plate.

Square wave

Exponential wave

1

2

3

4

5

6

7

8

9

10 11 12

A

200 V
2,000 μF
20 ms

250 V
2,000 μF
20 ms

300 V
2,000 μF
20 ms

250 V
350 μF
1000

Ω

250 V
500 μF
1000

Ω

250 V
750 μF
1000

Ω