Bio-Rad Gene Pulser MXcell™ Electroporation System User Manual
Page 42

33
Mixed protocols (page 38)
13. Opt 96 well/ Exp,Sqr
96
Use with cell lines with no protocol reference.
This protocol includes a range of average
starting conditions.
14. Opt 24 well/ Exp,Sqr
24
Use with cell lines with no protocol reference.
This protocol includes a range of average
starting conditions.
15. Opt 12 well/ Exp,Sqr
12
Use with cell lines with no protocol reference.
This protocol includes a range of average
starting conditions.
16. Uniform 96 well/ Exp,Sqr
96
Use with a set of defined conditions to
compare different cell lines and
electroporation of different molecules within
the same or different cell lines.
17. Uniform 24 well/ Exp,Sqr
24
Use with a set of defined conditions to
compare different cell lines and
electroporation of different molecules within
the same or different cell lines.
18. Uniform 12 well/ Exp,Sqr
12
Use with a set of defined conditions to
compare different cell lines and
electroporation of different molecules within
the same or different cell lines.
19. Mixed 96
96
Use for mixing different waveforms.
Alternating rows of exponential (250 V/350 uf)
and square waves (250 V/20 ms).
20. Mixed 24
24
Use for mixing different waveforms.
Alternating rows of exponential (250 V/350 uf)
and square waves (250 V/20 ms).
21. Mixed 12
12
Use for mixing different waveforms.
Alternating rows of exponential (250 V/350 uf)
and square waves (250 V/20 ms).
Table 6. List of pre-set protocols and their uses.
Protocol name
Number
of wells
When to use this protocol