TROUBLESHOOTING
59
TROUBLESHOOTING GUIDE
ERROR MESSAGES
Err1
Very Low battery. Replace battery or switch to AC power. Press back
arrow (
) to back out. Scan sample again. Replace battery as soon as
possible.
Err2
The meter can not be calibrated outside of the allowable range of the
displayed reading. Confirm that standard was made correctly. The
displayed reading can only be adjusted to ±50% of the factory
calibration.
Err3
Meter can not be calibrated with a zero sample. Calibrate the meter
with a sample other than zero.
Err4
Processing error due to motion of suspended particles or sub micron
air bubbles or opening/closing lid during readings. Scan sample again
until reading is obtained.
Err5
No blank reading. The meter has never been blanked for this test
factor. Blank meter.
Err6
Internal mathematical error. Re-blank the meter and rescan the
sample.
Err7
Configuration error. Call LaMotte Tech Service. Meter may have to be
returned for repairs.
low battery
Low battery. Change battery.
TROUBLESHOOTING
PROBLEM
REASON
SOLUTION
Lost in meter
menus
Reset to factory default
settings. See page 10.
Hold down
*IOK and press
ON. Release both buttons. Press
*IOK to select the default settings.
Meter will turn off and the factory
settings will be restored.
Unusually large
negative or
positive readings
when performing
calibration.
Incorrect standards used to
calibrate meter.
Use fresh 0.0 standard in clean,
labeled vial. Reset meter to factory
default settings. See procedure
above. Recalibrate meter.
Erroneous
readings
Measurement was taken with
lid open.
Close lid. Read again.
>
on display
Over range.
The sample is outside of the
acceptable range. Dilute sample
and test again.
Meter freezes
Lid was opened when
reading was being taken.
Close lid. Read again. Unplug
adapter. Plug adapter in to reset.