beautypg.com

LaMotte TC-3000 Tri-Meter User Manual

Page 55

background image

Calibration Procedure

The meter should be calibrated with platinum cobalt color standards. For the most
accurate results, a user calibration should be performed with LaMotte Color
Standards. The calibration should be done with a distilled or deionized water blank
and one color standard of known concentration. The concentration of the calibration
standard should be similar to the expected concentration of samples that will be
tested.

1. Press

ON to turn the meter on.

1 . 3

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Scroll down and then press *I

OK to

select Color.

M e a s u r e

T u r b i d i t y

* C o l o r

C h l o r i n e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

4. Rinse a clean tube (0290) with

color-free (distilled or deionized)
water. Fill the tube to the 10 mL
line with color-free water. Dry the
tube with a lint-free cloth. Put on a
dry positioning ring. Cap the tube.

5. Open the meter lid. Insert the tube

into the chamber. Align the index
notch on the positioning ring with
the index arrow on the meter. Close
the lid and press

*IOK to select

Scan Blank

.

C o l o r

* S c a n B l a n k

€

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

ANALYSIS & CALIBRATION • COLOR

55