13. Open the meter lid. Insert sample
into chamber. Align the index notch
on the positioning ring with the index
arrow on the meter. Close lid, select
*IOK to select Total Chlorine.
F r e e C h l o r i n e
( T )
1 . 0 0
p p m
* T o t a l C h l o r i n e
q
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
14. Record the result as Total Chlorine.
T o t a l C h l o r i n e
( T )
1 . 2 5
p p m
* S c a n S a m p l e
q
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
15. Subtract the Free Chlorine reading
from the Total Chlorine reading to
obtain the concentration of
Combined Chlorine.
16. Press
OFF to turn the meter off or
press
to exit to a previous menu or
make another menu selection.
*Warning: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD
or go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.
Note: For the most accurate results, samples over 6 ppm chlorine should be diluted
with chlorine demand free water and re-tested.
Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.
Total Chlorine
1. Press
ON to turn the meter on.
1 . 3
36
ANALYSIS & CALIBRATION • CHLORINE
Total
-
Free
= Combined
Chlorine
Chlorine
Chlorine