LaMotte TC-3000 Tri-Meter User Manual
Page 44
9. Add 5 drops of *DPD 3 Total
Chlorine Reagent (P-6741). Cap
and mix.
Note: For wastewater samples, Standard Methods for the Examination of Water and
Wastewater recommends waiting 2 minutes for full color development.
10. Open the meter lid. Insert the
sample into the chamber. Align the
index notch on the positioning ring
with the index arrow on the meter.
Close lid, select
*IOK to select
Scan Sample
.
T o t a l C h l o r i n e
( L )
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
11. Record the result as Total Chlorine.
T o t a l C h l o r i n e
( L )
1 . 0 0
p p m
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
12. Press
OFF to turn the meter off or
press
to exit to a previous menu
or make another menu selection.
*Warning: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD
or go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.
Note: For the most accurate results, samples over 6 ppm chlorine should be diluted
with chlorine demand free water and re-tested.
Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.
44
ANALYSIS & CALIBRATION • CHLORINE
DPD
3