beautypg.com

LaMotte TC-3000 Tri-Meter User Manual

Page 39

background image

LIQUID DPD REAGENTS
(Liquid calibration should be selected in the Options menu)

The letter (L) in the upper right corner of the display indicates that the meter is in liquid
mode.

Free Chlorine, Combined And Total Chlorine

1. Press

ON to turn the meter on.

1 . 3

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Scroll down and then press

*IOK

to select Chlorine.

M e a s u r e

T u r b i d i t y
C o l o r

* C h l o r i n e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

4. Press

*IOK to select Test Free

Chlorine.

C h l o r i n e

( L )

* T e s t F r e e

T e s t T o t a l

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

5. Rinse a clean tube (0290) with the

sample water. Fill the tube to the
10 mL line with the sample water.
Dry the tube with a lint-free cloth.
Put on a dry positioning ring. Cap
the tube.

ANALYSIS & CALIBRATION • CHLORINE

39