beautypg.com

LaMotte TC-3000 Tri-Meter User Manual

Page 52

background image

Testing Tips

1. Wash tubes thoroughly afer testing to prevent staining of tubes and contamination

of future test reactions with DPD residue.

2. When using liquid DPD reagents, invert bottle in a completely vertical position to

dispense uniform drops. Do not tilt bottle at an angle.

3. Follow instructions. Obey time limits.

4. In samples with extremely high chlorine concentrations, above 10 ppm, the

reagent system will show an initial flash of dark pink color that will fade quickly.
Dilute the sample and test again.

5. When testing salt water, double the amount of reagent used. Use ten drops of

each DPD liquid reagent or two DPD tablets.

6. Oxidized manganese (permanganate) will interfere with this test. Iodine and

bromine will give a positive interference.

7.

A permanganate check standard is not recommended for calibration when using
the liquid DPD reagent system.

8. The averaging option is not available for the chlorine test.

9. When testing at low concentrations use the same tube for the blank and the

sample.

10. Always use the tube positioning ring. Always insert tube into the meter chamber

with the same amount of pressure and to the same depth.

11. Occasionally clean the chamber with a damp lint-free wipe, followed by an

alcohol dampened wipe. A clean cahmber and tubes are essential for reliable
results.

COLOR

Test results are reported as cu (Color Units).

Analysis

1. Press

ON to turn the meter on.

1 . 3

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

52

ANALYSIS & CALIBRATION • CHLORINE • COLOR