LaMotte TC-3000 Tri-Meter User Manual

Page 12

background image

2. Scroll down and then press

*IOK to

select Options.

M a i n M e n u

M e a s u r e
D a t a L o g g i n g

* O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

3. Scroll down and then press

*IOK

to select Turbidity.

O p t i o n s

A v e r a g i n g

* T u r b i d i t y

C h l o r i n e
D a t e / T i m e

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

4. Scroll down and then press

*IOK

to select Calibration.

T u r b i d i t y

U n i t s

* C a l i b r a t i o n

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

5. Scroll down and then press

*IOK

to select a Calibration curve. Select
a calibration option based on the
composition of the standards that
will be used to calibrate the meter.
Available options are: Formazin,
Polystyrene

, Attenuation.

T u r b i d i t y

U n i t s

* C a l i b r a t i o n

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

For the most accurate results, the
Attenuation option should be chosen
when samples are over 500 NTU.
The range for the attenuation option
is 40–4000 NTU.

C a l i b r a t i o n
* F o r m a z i n

J a p a n S t a n d a r d
A t t e n u a t i o n

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

Note: StablCal

â

standards below 50 NTU should not be used to calibrate the

TC-3000. The diluent has a different refractive index than traditional formazin
standards and will affect the results. The Japan Standard calibration mode should
be used only with Japanese Polystyrene Standards (0–100 NTU).

12

OPTIONS & SETUP • TURBIDITY