beautypg.com

LaMotte TC-3000 Tri-Meter User Manual

Page 51

background image

C a l i b r a t e

0 1 . 3 5

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

C a l i b r a t e

0 1 . 3 0

q, Q, p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

12. When the value on the display

matches the concentration of the
chlorine standard, press

*IOK to

select Set.

C a l i b r a t e

1 . 3 0

* S e t

‚

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

Or press the

ˆ and press *IOK to

return the meter to the default
setting.

C a l i b r a t e

1 . 3 0

* D e f a u l t

p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

13. Press

*IOK to proceed to Chlorine

analysis. Press

OFF to turn the meter

off or press

… to exit to a previous

menu or make another menu
selection.

*Warning: Reagents marked with an * are considered to be potential health hazards.
To view or print a Material Safety Data Sheet (MSDS) for these reagents see MSDS CD
or go to www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail,
phone or fax.

Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.

ANALYSIS & CALIBRATION • CHLORINE

51

3

0