beautypg.com

LaMotte TC-3000 Tri-Meter User Manual

Page 33

background image

11. Excessive color in a sample will absorb light and cause high readings. The user

should verify if a certain level of color will cause a significant error at the level of
turbidity being tested.

12. Observe shelf life recommendations for turbidity standards.

13. Do not use silicone oil on tubes when testing turbidity with the TC-3000.

14. When testing at low concentrations use the same tube for the blank and the

sample.

15. Always use the positioning ring. Always insert tube into the meter chamber with the

same amount of pressure and to the same depth.

16. Occasionally clean the chamber with a damp lint-free wipe, followed by an

alcohol dampened wipe. A clean chamber and tubes are essential for reliable
results.

17. For the greatest accuracy during the calibration procedure, be sure that after the

meter is blanked and the blank is scanned as a sample, the reading is 0.00. If
not, reblank the meter and scan the blank again until it reads 0.00. When
scanning calibration standard as the sample, scan the calibration standard three
times removing the tube from the chamber after each scan. The readings should
be consistent. Use the last consistent reading to calibrate the meter. If the readings
are not consistent, avoid using an aberrant reading to calibrate the meter.

CHLORINE

The default units are ppm and the default calibration curve is for DPD Tablet reagents.
For the most accurate results, a user calibration should be performed. The letter (T) in
the upper right corner of the display indicates that the meter is in the tablet mode. To
use liquid DPD reagents, see the Set Up instructions.

Analysis

TABLET DPD REAGENTS

Free Chlorine, Combined And Total Chlorine

1. Press

ON to turn the meter on.

1 . 3

2. Press

*IOK to select Measure.

M a i n M e n u
* M e a s u r e

D a t a L o g g i n g
O p t i o n s

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

ANALYSIS & CALIBRATION • CHLORINE

33