beautypg.com

LaMotte TC-3000 Tri-Meter User Manual

Page 32

background image

17. When the value on the display

matches the concentration of the
turbidity standard, press the

*IOK

to select Set.

C a l i b r a t e

0 0 . 5 0

* S e t

q

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

Or press

ˆ press *IOK to return

the meter to the default setting.

C a l i b r a t e

0 0 . 5 0

* D e f a u l t

p

1 6 : 0 2 : 1 9

0 1 / 0 4 / 0 5

18. Press

*IOK to proceed to Turbidity

analysis. Press

OFF to turn the meter

off or press

… to exit to a previous

menu or make another menu
selection.

Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.

Testing Tips

1. Samples should be collected in a clean glass or polyethylene container.

2. Samples should be analyzed as soon as possible after collection.

3. Gently mix sample by inverting before taking a reading but avoid introducing air

bubbles.

4. For the most precise results, follow the recommended procedure for wiping a filled

tube before placing it in the meter chamber. Invert tube very slowly and gently
three times to mix the sample. Surround the tube with a clean, lint-free cloth. Press
the cloth around the tube. Rotate the tube in the cloth three times to assure that all
areas of the tube have been wiped.

5. Discard tubes that have significant scratches and imperfections in the light pass

zones. (Central zone between bottom and fill line).

6. When reading very low turbidity samples, do not use tubes or caps that have been

used previously with high turbidity samples.

7.

Use the averaging option for low level measurements of turbidity.

8. The meter should be placed on a surface that is free from vibrations. Vibrations

can cause high readings.

9. Turbidity readings will be affected by electric fields around motors.

10. Carbon in the sample will absorb light and cause low readings.

32

ANALYSIS & CALIBRATION • TURBIDITY