LaMotte TC-3000 Tri-Meter User Manual
Page 38
8. Add one *Chlorine DPD #3
Instrument Grade Tablet (6197).
Crush tablet with Tablet Crusher
(0175). Cap tube and shake until
the tablet dissolves.
Note: For wastewater samples, Standard Methods for the Examination of Water and
Wastewater recommends waiting 2 minutes for full color development.
9. Open the meter lid. Insert sample
into chamber. Align the index notch
on the positioning ring with the index
arrow on the meter. Close lid, select
*IOK to select Scan Sample.
T o t a l C h l o r i n e
( T )
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
10. Record the result as Total Chlorine.
T o t a l C h l o r i n e
( T )
1 . 0 0
p p m
* S c a n S a m p l e
1 6 : 0 2 : 1 9
0 1 / 0 4 / 0 5
11. Press
OFF to turn the meter off or
press
to exit to a previous menu
or make another menu selection.
*Warning: Reagents marked with an * are potential health hazards
.
To view or print a
Material Safety Data Sheet (MSDS) for these reagents see MSDS CD or go to
www.lamotte.com. To obtain a printed copy, contact LaMotte by e-mail phone or fax.
Note: For the most accurate results, samples over 6 ppm chlorine should be diluted
with chlorine demand free water and re-tested.
Note: The meter will remember the last scanned blank reading. It is not necessary to
scan a blank each time the test is performed. To use the previous blank reading,
instead of scanning a new one, scroll to Scan Sample and proceed. For the most
accurate results, the meter should be blanked before each test and the same tube with
tube positioning ring should be used for the blank and the reacted sample.
38
ANALYSIS & CALIBRATION • CHLORINE