beautypg.com

Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual

Page 85

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

■■■■■■■■

75

NOTE

The maximum treatment time in an ultrasonic bath is 20 s. Ultra-
sonic baths that last longer than this can damage the ruby ball
inside the valve.

The outlet valve has to be replaced if it remains clogged after the
ultrasonic bath.

3 Reinserting the outlet valve into the pump head

Insert the outlet valve into the outlet valve holder (the seal must
be visible) (1).

Screw the outlet valve holder up into the pump head (2) and
tighten it in place using the adjustable wrench (3).

Tighten the connection capillary to the auxiliary piston back onto
the outlet valve holder.

Servicing the inlet valve

Accessories

For this step you need the following accessories:

You can find these parts in the accessory kit: Vario/Flex Basic
(6.5000.000).

Adjustable wrench (6.2621.000) from the accessory kit: Vario/Flex
Basic
(6.5000.000)