3 Installation

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

■■■■■■■■

41

Install the pump tubing connection with safety device and filter
(6.2744.180) at the pump tubing outlet:

– Push the locknut onto the pump tubing.
– Push the appropriate adapter onto the pump tubing.
– Place the tubing olive with the filter holder into the pump

tubing so that the pump tubing is firmly in place; push the
end of the pump tubing over at least the second notch of
the olive.

– Tighten it using the union nut.

2 Removing the tubing cartridge

Press in the tubing cartridge's snap-action lever.

Tilt the tubing cartridge upwards.

Unhook the tubing cartridge from the mounting bolt.

3 Inserting the pump tubing

Press the tubing cartridge's contact pressure lever down all the
way.

Place the pump tubing in the tubing cartridge. Fit the tubing car-
tridge between two stoppers. The stoppers must snap into the
corresponding holder of the tubing cartridge.

4 Inserting the tubing cartridge

Hang the tubing cartridge in the mounting bolt and press it in the
cartridge holder until you hear the snap-action lever snap in.

Connecting capillaries for regeneration solution

Accessories

For this step you need the following accessories:

Aspiration capillary (6.1803.020)

2 × pressure screw, short (6.2744.070)

On the pump tubing's …

… connect this accessory