beautypg.com

Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual

Page 105

background image

■■■■■■■■■■■■■■■■■■■■■■

5 Operation and maintenance

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

■■■■■■■■

95

The rotor's sealing surface is approximately 4 mm inside the hous-
ing if the rotor is inserted correctly. If this is not the case, the
rotor must be moved into the right position from below using a
sharp object.

5 Cleaning the new connecting piece

Clean the surface of the new connecting piece (25-2) with etha-
nol using a lint-free cloth.

6 Inserting the new connecting piece

Insert the connecting piece (25-2) into the housing (25-4) in such
a way that the connector 1 is on top and the three pins of the
connecting piece fit into the corresponding recesses on the hous-
ing.

Reattach the union nut (25-1) and tighten it by hand.

7 Connecting and conditioning the suppressor

Reconnect all capillaries of the suppressor to the IC system.

Before switching the suppressor over for the first time, rinse the
three suppressor units with solution for 5 minutes.