beautypg.com

3 connecting the rinsing solution – Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual

Page 46

background image

3.13 Suppressor

■■■■■■■■■■■■■■■■■■■■■■

36

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

1 Prepare a tubing cartridge of the peristaltic pump for the regenera-

tion solution (see Chapter 3.14.1, page 37).

2 Fasten the capillary labeled regenerant to the outlet of the pump

tubing using a pressure screw (6.2744.070).

3 Fasten the PTFE capillary from the bottle with the regeneration solu-

tion to the inlet of the pump tubing.

3.13.2.3

Connecting the rinsing solution

There are 2 options for rinsing the suppressor:

Rinsing solution via STREAM (recommended)
Use the eluent from the conductivity detector as a rinsing solution.

Rinsing solution via peristaltic pump
Prepare the rinsing solution in its own bottle and transport with the
peristaltic pump.

The rinsing solution is connected to the rinsing solution capillary.

Connecting the rinsing solution inlet with STREAM

1 Connect the detector output capillary and the capillary labeled rins-

ing solution to each other using a coupling (6.2744.040) and two
pressure screws (6.2744.070).