3 Installation

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

■■■■■■■■

25

3.8

Connecting the eluent degasser

The high-pressure pump cannot generate uniform flow if the eluent con-
tains small gas bubbles or dissolved gas. As a result, the baseline cannot
be stabilized correctly. In order to achieve good measurement results, the
eluent has to be degassed before it goes into the high-pressure pump.

The eluent degasser is completely connected. No installation work is
required.

3.9

Installing the high-pressure pump

The intelligent and low-pulsation high-pressure pump pumps the eluent
through the system. It is equipped with a chip where its technical specifi-
cations and "life history" (operating hours, service data, etc.) are saved.

The high-pressure pump consists of:

The pump head, which pumps the eluent through the system.

The purge valve used for bleeding the pump head.