3.19 Connecting the instrument to a computer

■■■■■■■■■■■■■■■■■■■■■■

52

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

4 Connecting the rinsing solution output capillary

Fasten the capillary labeled waste rins. to the waste collector. You
can find information on installing the waste collector in the leaflet
accompanying the IC equipment: waste collector (6.5336.000).

Alternative: Direct the capillary labeled waste rins. into a suffi-
ciently sized waste container and fasten it in place there.

5 Connecting the regeneration solution input capillary

Fasten the capillary labeled regenerant to the pump tubing con-
nection carrying the regeneration solution using a short pressure
screw (6.2744.070).

6 Connecting the regeneration solution output capillary

Fasten the capillary labeled waste reg. to the waste collector
(6.5336.000). You can find information on installing the waste
collector in the leaflet accompanying the IC equipment: waste
collector (6.5336.000).

Alternative: Direct the capillary labeled waste reg. into a suffi-
ciently sized waste container and fasten it in place there.

3.19

Connecting the instrument to a computer

NOTE

The instrument must be switched off when being connected to a com-
puter.

Accessories

For this step you need the following accessories:

USB connecting cable (6.2151.020)