3 installing adsorption cartridges, Installing adsorption cartridges – Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual

Page 56

background image

3.15 Metrohm CO

2

Suppressor (MCS)

■■■■■■■■■■■■■■■■■■■■■■

46

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

3 Connection to the detector

Connect the detector input capillary with a long pressure screw
(6.2744.090) to the MCS output (labeled Out).

CAUTION

If the MCS is not used, the input and output must be sealed with the
threaded stoppers (6.2744.220).

3.15.3

Installing adsorption cartridges

The suctioned air has to have the lowest CO

2

content possible in order for

the CO

2

to be removed from the eluent as efficiently as possible. In order

to achieve this, the air is aspirated through a CO

2

adsorption cartridge

(6.2837.000).

However, the CO

2

adsorption cartridge is sensitive to humidity; the CO

2

adsorption cartridge becomes clogged when wet. In order to protect the
CO

2

adsorption cartridge, an H

2

O adsorption cartridge (6.2837.010) is

also installed.

Accessories

For this step you need the following accessories:

Adsorption cartridge holder (6.2057.080) with 4 clips, an adapter
(6.1808.190) and a piece of PVC tubing.

CO

2

adsorption cartridge (6.2837.000)

H

2

O adsorption cartridge (6.2837.010)

All of the parts are located in one accessory kit: Vario/Flex SeS
(6.5000.020).

Installing the adsorption cartridge

CAUTION

The following preparatory steps absolutely must be carried out for CO

2

suppression to operate correctly.

Preparing the adsorption cartridges:

Prepare the H

2

O adsorption cartridge for use as specified in the provi-

ded leaflet.

Remove the protective cap on the tip of the CO

2

adsorption cartridge.