beautypg.com

2 intended use, Intended use – Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual

Page 13

background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

■■■■■■■■

3

Peristaltic pump
The peristaltic pump is used for pumping sample and auxiliary solutions. It
can rotate in both directions.

Metrohm CO

2

Suppressor (MCS)

The Metrohm CO

2

Suppressor (MCS) removes the CO

2

from the eluent

flow. This lowers the background conductivity, improves detection sensi-
tivity and minimizes the injection peak and carbonate peak.

Detector
Metrohm offers a series of different detectors for various analysis tasks. A
suitable detector type must be ordered as a separate accessory.

Sample degasser
The sample degasser removes gas bubbles and dissolved gases from the
sample.

Sample preparation module (SPM)
The SPM is used for inline sample preparation. Its important applications
include neutralizing samples and cation exchange before sample injection.

Separation column
The intelligent separation column separates different components accord-
ing to their interactions with the column. Metrohm separation columns
are equipped with a chip where their technical specifications and history
(start-up, operating hours, etc) are stored.

1.2

Intended use

The 940 Professional IC Vario ONE/SeS/PP/Prep 3 is used for the determi-
nation of anions or polar substances with sequential suppression using ion
chromatography.

Sequential suppression consists of:

Chemical suppression with a Metrohm Suppressor Module (MSM) and
subsequent

CO2 suppression with the Metrohm CO2 Suppressor (MCS).

Background conductivity is reduced to a minimum with sequential sup-
pression.

The instrument can also be used as needed for the determination of cati-
ons or anions without chemical suppression.

The present instrument is suitable for processing chemicals and flammable
samples. Usage of the 940 Professional IC Vario therefore requires the