Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual
Page 68
3.22 Connecting and rinsing the guard column
■■■■■■■■■■■■■■■■■■■■■■
58
■■■■■■■■
940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)
■
Guard column (suitable for the separation column)
Connecting the guard column
1 Removing the coupling
Remove the coupling installed between the column input capillary
and the column output capillary for the initial start-up.
2 Preparing the guard column
■
Remove the stopper and the sealing cap from the guard column.
3 Connecting the guard column
CAUTION
When inserting the guard column, ensure that it is inserted cor-
rectly based on the marked flow direction (if specified).
■
Fasten the input of the guard column to the column input capil-
lary using a short pressure screw (6.2744.070).
■
Optional: In case the guard column is connected to the separation
column using a provided a connection capillary, fasten the con-
nection capillary to the guard column's output using the pressure
screw also provided.
Rinsing the guard column
1 Rinsing the guard column
■
Place a beaker under the guard column's output.
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)