3 Installation

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

■■■■■■■■

61

Connecting the separation column

1 Removing the stoppers

CAUTION

When inserting the column, ensure that it is inserted correctly
based on the marked flow direction.

Remove the stoppers from the separation column.

2 Installing the input of the separation column

There are 3 possibilities:

Attaching the bottom end of the separation column directly to
the guard column.
or

If the guard column is connected to the separation column using
a provided connection capillary, connect the bottom end of the
separation column to the guard column's output capillary using
the provided PEEK pressure screw (6.2744.070).
or

If no guard column is used (not recommended), connect the col-
umn input capillary to the input of the separation column using a
short pressure screw (6.2744.070).

3 Rinsing the separation column

Place a beaker under the separation column's outlet.

Start manual control in MagIC Net and select the high-pressure
pump: Manual

Manual control Pump

Flow: in accordance with column leaflet
On

Rinse the separation column with eluent for approximately 10
minutes.

Use manual control in MagIC Net to stop the high-pressure pump
again: Off.

4 Removing the coupling

Remove the coupling (6.2744.040) from the column output capil-
lary.