beautypg.com

7 connecting the eluent bottle, Connecting the eluent bottle, Figure 7 – Metrohm 940 Professional IC Vario ONE/SeS/PP/Prep 3 User Manual

Page 32: Installing the eluent bottle cap

background image

3.7 Connecting the eluent bottle

■■■■■■■■■■■■■■■■■■■■■■

22

■■■■■■■■

940 Professional IC Vario ONE/SeS/PP/Prep 3 (2.940.1530)

3.7

Connecting the eluent bottle

The eluent is aspirated out of the eluent bottle via the eluent aspiration
tubing . The eluent aspiration tubing is installed on the input for the elu-
ent degasser.

The tubing has to be be fed out of the instrument through a suitable
opening (see "Openings on the back panel", page 12) before the loose
end can be connected to the eluent bottle.

Accessories

For this step you need the following accessories:

These parts are part of the accessory kit Vario/Flex ONE (6.5000.010).

Eluent bottle (6.1608.070)

The eluent bottle cap GL 45 accessory set (6.1602.160)
This accessory set contains the bottle cap, an M6 tubing nipple, an M8
tubing nipple, two O-rings and an M6 and M8 threaded stopper.

The tubing adapter for aspiration filter accessory set (6.2744.210)
This accessory set contains a filter holder, a clamping screw and tubing
weighting.

An aspiration filter (6.2821.090)

Connecting the eluent aspiration tubing

1 Installing the eluent bottle cap (6.1602.160)

Start by pushing the M8 tubing nipple onto the loose end of the
eluent aspiration tubing, followed by the O-ring.

Push the loose end of the eluent aspiration tubing through the
M8 opening of the bottle cap and screw it on for the time being.

Figure 7

Installing the eluent bottle cap