5 Appendix

Mains connection

Voltage

115 V: 100...120 V ± 10 %

230 V: 220...240 V ± 10 %

Switchable with voltage selector insert in the fuse

holder (see section 2.2.1)

Frequency

50...60 Hz

Power consumption

5 VA

Fuse

5 mm dia., 20 mm length

100…240 V: 1 A (slow-blow)

Safety specifications

Construction/testing

According to IEC 1010 / EN 61010 / UL 3101-1,

protection class 1, degree of protection IP40

Safety directions

The Instructions for Use include information and

warnings which must be heeded by the user to

assure safe operation of the instrument.

Electromagnetic compatibility (EMC)

Emitted interference

Standards met:

EN55011 (class B), EN55022 (class B),

EN 50081-1/2

Immunity to interference

Standards met:

IEC801-2/IEC1000-4-2 (class 3), IEC801-4/

IEC1000-4-4 (class 3), IEC1000-3/EN61000-3,

IEC1000-4-11/EN61000-4-11

Ambient temperature

Nominal operating range

+5…+40°C

(at 20…80 % atmospheric humidity)

Storage, transport

–40…+70°C

Housing

Material of cover

Polyurethane rigid foam (PUR) with fire protection

for fire class UL94VO, CFC-free

Material of base

Steel, enameled

Dimensions

Width

260 mm

Height

129 mm

Depth

366 mm

Weight

3.6 kg (with accessories)

791 VA Detector