beautypg.com

Metrohm 791 VA Detector User Manual

Page 34

background image

3 Operation

791 VA Detector

28

3 Overload test

For this test the working electrode must not be connected to the

791 VA Detector:

Set the polarization voltage on the 791 VA Detector to the

required value for the following determinations with digital

switch 11.

Set commutator 3 on the 791 VA Detector to "meas". Over-

load display 2 must light up briefly and then go out again.

If overload display 2 remains lit up for a longer period of time

this indicates that there is a faulty electrical contact between

the reference electrode and the auxiliary electrode. Eliminate

the cause (e.g. air in the system or in the reference electrode,

blocked diaphragm, faulty cable connection) and repeat the

test.

If the overload test is OK, set commutator 3 on the 791 VA

Detector back to "Stand-by".

4 Equilibrating the working electrode

Connect working electrode 28 to connection 18 of the 791 VA

Detector with the 6.2120.020 Cable.

Set commutator 3 on the 791 VA Detector to "meas".

Observe equilibration curve: the curve should flatten out

noticeably within a relative short time (10

20 min), otherwise

the working electrode must be repolished or refilled.

In current measuring ranges > 5 nA a constant base current

is usually achieved with one hour. At higher sensitivities the

working electrode should be equilibrated overnight.

5 Zero balancing the base current

(only for recorders)

Set the current measuring sensitivity on the 791 VA Detector

with knob 8 to the required value.

Use potentiometer 9 on the 791 VA Detector to adjust the

base current to the zero point of the recorder. If the setting

range of potentiometer 9 is not sufficient then it can be ex-

tended by using commutator 5 (

×

5) or 6 (

Ч

50).