4 Maintenance – Malfunctions

791 VA Detector

36

4 Set selector 20 on the rear panel to .

5 Set selector 21 on the rear panel to 0

1 V.

0...10 mV

0...1 V

6 Set recorder sensitivity to 1000 mV.

7 Set commutator 3 "meas" to "on" position

(depressed position).

meas

8 Set potentiometer 9 I

comp

abs. to 0.

0.02 nA

Icomp

9 Bring recorder of voltmeter to 0 using potenti-

ometer 10 I

comp

rel..

Icomp

. I

0.01

10 Alter the settings of digital switch 11 U

pol

in

steps of 100 mV and 10 mV: The settings

shown on the digital switch should be repro-

duced on the recorder or voltmeter within a

tolerance of <12.5 mV and <1.25 mV respec-

tively.

1

4

0

+

11 Set digital switch 11 U

pol

to and alter the

settings in steps of 100 mV and 10 mV: The

settings shown on the digital switch should be

reproduced on the recorder or voltmeter within

a tolerance of <12.5 mV and <1.25 mV respec-

tively.

2

5

0


4.4.3

Check of current compensation

1 Set digital switch 11 U

pol

to 0 mV.

0

0

0

+

2 Set adjusting knob 8 to 0.1

µ

A.

4

5

1

10

50

1

5

0,1

0,5

100
10

0,5

0,1