5 check sensitivity knob, 5 check sensitivity knob 8 – Metrohm 791 VA Detector User Manual

Page 45

background image

4.4 Instrument test with the dummy cell

791 VA Detector

39

11 Set commutator 3 "meas" to "on" position

(depressed position).

meas

12 Unplug dummy cell at "AE": the overload

display 2 lights up.




4.4.5

Check sensitivity knob 8

1 Unplug dummy cell.

2 Set adjusting knob 8 to 50 nA.

4

5

1

10

50

1

5

0,1

0,5

100
10

0,5

0,1

3 Set commutators 3

7 to "off" position (non-

depressed position).

4 Set potentiometer 9 I

comp

abs. to 0.

0.02 nA

Icomp

5 Set selector 21 on the rear panel to 0

1 V.

0...10 mV

0...1 V

6 Bring recorder of voltmeter to 0 using potenti-

ometer 10 I

comp

rel. (set line recorder first to mid-

scale).

Icomp

. I

0.01

7 Turn adjusting knob 8 through all positions and

check the deviation from zero after a waiting

time of app. 5 s. Deviation should be <

±

1 V.