beautypg.com

5 practical information about elcd, Cleanliness, Pulsation – Metrohm 791 VA Detector User Manual

Page 24: Eluent

background image

3 Operation

791 VA Detector

18

If no signal is observed, change the pH value of the eluent (a proto-

nated or deprotonated substance may be oxidized more easily). Gen-

erally, a higher polarization voltage has to be applied with lower pH val-

ues (approx. 50

100 mV per pH unit).

1

st

experiment with eluent of pH

6

2

nd

experiment with eluent of pH

3

If there is still no signal, the analyte is "inactive", i.e. not suitable for elec-

trochemical detection.
As soon as a signal is observed, decrease the polarization voltage in

steps of 100 mV: a hydrodynamic current/voltage curve results.
For detection, select the lowest polarization voltage U

pol

at which a

good signal is still observed.


3.1.5

Practical information about ELCD

In principle the electrochemical detector can be connected to all HPLC

and IC systems. In order to carry out work successfully, particularly in

trace-level quantities, all components such as pump, dampening sys-

tem, injector, separating column, capillaries, couplings, etc. must be in

perfect condition. In electrochemical detection it is possible that inter-

ferences may occur which, for example, are not observed with UV or

conductivity detection. These can often be traced back to contaminants

from the water supply system (e.g. Fe(II) ions), contaminants from the

separating column (e.g. organic residues) or variations in pressure

caused by leaking connections.
Particular attention must be given to the following points:

Cleanliness

The general demands placed on trace analysis work also apply here.

Each possible source of corrosion of steel components must be

avoided as Fe(II) ions may interfere. The eluent should also not contain

any complexing agents.

Pulsation

The noise of the detector signal depends on the pressure variation in-

side the cell and on the back-pressure (max. 1 bar), which is deter-

mined by the length of outlet capillary 38 (i.d. 0.3 mm). The use of a

pulsation dampener (e.g. Metrohm 6.2620.150 Pulsation dampener) for

dampening such pulsations has proved itself in practice and is also

adequate if only single-piston pumps are used.

Eluent

Most of the solvents normally used in HPLC and IC can be used to pre-

pare eluents. The conductivity which is necessary for detection is

achieved by the addition of a conductive salt. Concentrations of 1

10

g/L are normally adequate for this purpose, this results in a conductivity

of approx. 1

10 mS/cm. Sulfates, nitrates, phosphates, acetic acid,