Metrohm 791 VA Detector User Manual
Page 39

4.3 Malfunctions and their rectification
791 VA Detector
33
Malfunction
Cause
Rectification
Noise level too high
•
Inadequate pulsation
damping.
•
Electrostatic screening
defective.
•
Background current too high.
•
Eluent outlet tubing under
electrostatic influence.
•
Defective working electrode.
•
Defective 791 VA Detector.
•
Use pulsation dampener.
•
Check screening, if necessary
connect steel column and/or its
outlet to socket 14 of the VA
Detector by means of a cable.
•
Check screening, if necessary
connect steel column and/or its
outlet to socket 14 of the VA
Detector by means of a cable.
•
Immerse eluent outlet tubing
completely in spent solution in
waste recipient or renew PTFE
tubing if necessary.
•
Replace working electrode.
•
Check 791 VA Detector using
the dummy cell (see section
4.1.4)
Spikes on baseline
•
Air in detector cell.
•
Air entering detector cell.
•
If a CPE working electrode is
in use, the carbon paste is not
in good condition.
•
Unscrew working electrode and
deaerate detector cell.
•
Degas eluent or subject it to
appropriate treatment.
•
Refill CPE and ensure paste is
well tamped down.
Isolated spikes on
baseline
•
Electrostatic discharges from
operating personnel on the
apparatus.
•
Mains interference.
•
Electromagnetic interference
from environment.
•
Such discharges cannot be
completely avoided, but are not
usually too troublesome at
normal (i.e. not too dry) atmos-
pheric humidity.
•
Connect all instruments to the
power supply via a suitable filter
(e.g. Metrohm 615 Mains Dis-
tributor).
•
Change "environment".
Substance turnover
too low or non-
existent
•
Polarization voltage too low.
•
Active surface of working
electrode blocked by sub-
stances.
•
Increase polarization voltage.
•
Polish working electrode or
renew paste in CPE.
Substance turnover
falls off
•
Active surface of working
electrode blocked by sub-
stances.
•
Quantity of substance too
great.
•
Polish working electrode or
renew paste in CPE.
•
The sample quantity should be
less than 100 ng per injection to
avoid blocking of the electrode
surface.
Front peak too high
•
Environment altered by
sample solution.
•
Eluent contaminated by Fe(II)
ions from rusty steel parts
(traces only suffice!).
•
Whenever possible, use eluent
as solvent for the sample solu-
tion.
•
Check entire HPLC or IC system
for rust; check resistance of all
materials to the reagents in use.