beautypg.com

Metrohm 791 VA Detector User Manual

Page 39

background image

4.3 Malfunctions and their rectification

791 VA Detector

33

Malfunction

Cause

Rectification

Noise level too high

Inadequate pulsation
damping.

Electrostatic screening
defective.

Background current too high.


Eluent outlet tubing under
electrostatic influence.

Defective working electrode.

Defective 791 VA Detector.

Use pulsation dampener.

Check screening, if necessary
connect steel column and/or its
outlet to socket 14 of the VA
Detector by means of a cable.

Check screening, if necessary
connect steel column and/or its
outlet to socket 14 of the VA
Detector by means of a cable.

Immerse eluent outlet tubing
completely in spent solution in
waste recipient or renew PTFE
tubing if necessary.

Replace working electrode.

Check 791 VA Detector using
the dummy cell (see section
4.1.4
)

Spikes on baseline

Air in detector cell.

Air entering detector cell.

If a CPE working electrode is
in use, the carbon paste is not
in good condition.

Unscrew working electrode and
deaerate detector cell.

Degas eluent or subject it to
appropriate treatment.

Refill CPE and ensure paste is
well tamped down.

Isolated spikes on
baseline

Electrostatic discharges from
operating personnel on the
apparatus.

Mains interference.


Electromagnetic interference
from environment.

Such discharges cannot be
completely avoided, but are not
usually too troublesome at
normal (i.e. not too dry) atmos-
pheric humidity.

Connect all instruments to the
power supply via a suitable filter
(e.g. Metrohm 615 Mains Dis-
tributor).

Change "environment".

Substance turnover
too low or non-
existent

Polarization voltage too low.

Active surface of working
electrode blocked by sub-
stances.

Increase polarization voltage.

Polish working electrode or
renew paste in CPE.

Substance turnover
falls off

Active surface of working
electrode blocked by sub-
stances.

Quantity of substance too
great.

Polish working electrode or
renew paste in CPE.

The sample quantity should be
less than 100 ng per injection to
avoid blocking of the electrode
surface.

Front peak too high

Environment altered by
sample solution.

Eluent contaminated by Fe(II)
ions from rusty steel parts
(traces only suffice!).

Whenever possible, use eluent
as solvent for the sample solu-
tion.

Check entire HPLC or IC system
for rust; check resistance of all
materials to the reagents in use.