3.4.3 Measuring

procedure

In order to be able to carry out measurements it is essential that the

working electrode has achieved a stable base current (see section

3.4.2). If the value of the base current regains its original value after the

determination then as many determinations as required can be carried

out. Attention should be given to the following points:

1 Carrying out the determination

Inject the sample solution into the HPLC or IC system.

Record and evaluate the current curve.

2 Allow ELCD system to run overnight

If the ELCD system is to be used again on the following day

then the 791 VA Detector should not be switched off over-

night. The working electrode is constantly under the polariza-

tion current and is quickly ready for new measurements.

The HPLC or IC pump is set to a minimal flow rate and runs

overnight. This means that the base current is stabilized after

only a few minutes.

3 Switching off the ELCD system

Set commutator 3 on the 791 VA Detector to "Stand-by".

Unplug working electrode 28.

Switch off the 791 VA Detector.