4 check of current and voltage overload – Metrohm 791 VA Detector User Manual

Page 44

background image

4 Maintenance – Malfunctions

791 VA Detector

38

4.4.4

Check of current and voltage overload

1 Set digital switch 11 U

pol

to +10 mV or

–10 mV.

0

0

1

+

2 Set adjusting knob 8 to 50 nA.

4

5

1

10

50

1

5

0,1

0,5

100
10

0,5

0,1

3 Set selector 20 on the rear panel to .

4 Set selector 21 on the rear panel to 0

1 V.

0...10 mV

0...1 V

5 Set potentiometer 9 I

comp

abs. to 0.

0.02 nA

Icomp

6 Set commutators 3

7 to "off" position (non-

depressed position).

7 Bring recorder of voltmeter to 0 using potenti-

ometer 10 I

comp

rel..

Icomp

. I

0.01

8 Set commutator 6 "

×

50" to "on" position

(depressed position).

x50

9 Turn potentiometer 9 I

comp

abs. clockwise until

overload display 2 lights up: The voltage at the

recorder output should be app. 10.5

12 V.

Without digital voltmeter, read the value on the

scale of potentiometer 9: The value should be

between 500 and 620 scale divisions.

0.02 nA

Icomp

10 Reset potentiometer 9 I

comp

abs. to 0: The

overload display 2 extinguishes.

0.02 nA

Icomp