beautypg.com

3 malfunctions and their rectification – Metrohm 791 VA Detector User Manual

Page 38

background image

4 Maintenance – Malfunctions

791 VA Detector

32

4.3

Malfunctions and their rectification

Malfunction

Cause

Rectification

No signal

VA Detector is in "stand-by"
mode.

VA Detector switched off or
disconnected from mains
voltage.

Recorder switched off or
disconnected from mains
voltage.

Recorder disconnected from
VA Detector.

Electrodes disconnected.

Broken interconnecting
cables.

Press "meas" button on VA
Detector.

Connect VA Detector and
switch on.

Connect recorder and switch
on.

Connect recorder to the VA
Detector.

Connect electrodes.

Use new interconnecting
cables.

VA Detector
shows "overload"

Current range set exceeded
by too high a concentration of
substance.

The VA Detector is either
incorrectly balanced or not
balanced at all.

Reference electrode discon-
nected.

Auxiliary electrode discon-
nected.

Polarization voltage badly
adjusted.

Wait or set a higher current
range at the VA Detector.

Balance the recording instru-
ment and VA Detector correctly.

Connect reference electrode.

Connect auxiliary electrode.

Adjust polarization voltage
correctly.

Excessively high
background current

Polarization voltage too high.

Eluent contaminated.

Eluent contaminated by Fe(II)
ions from rusty steel parts
(traces only suffice!)

Polarization voltage should
generally not be set higher than
1200 mV.

Prepare fresh eluent using only
the purest (AR grad) chemicals.

Check entire HPLC or IC system
for rust; check resistance of all
materials to the reagents in use.

Baseline unstable

Highly retarded or enriched
components from the separat-
ing column.

Pressure fluctuations due to
poor seals, blocked filters or
capillaries, defective pump,
etc.

Eluent contaminated by Fe(II)
ions from rusty steel parts
(traces only suffice!).

Defective reference electrode.

Rinse the column well with
eluent or regenerate it, use new
column if necessary.

Check HPLC or IC system for
these faults.

Check entire HPLC or IC system
for rust; check resistance of all
materials to the reagents in use.

Replace reference electrode.