3.2 Electrodes

791 VA Detector

21

Carbon paste electrode CPE

Ordering designation 6.0807.000
Sensor material

paste composed of spectroscopic grade car-

bon powder and paraffin

Chemical resistance

good resistance to aqueous solutions; limited

resistance only to water/acetonitrile and wa-

ter/methanol mixtures

Voltage range (–200)

0

+1200 mV (pH < 7)

Regeneration

The sensor surface of the CPE is more vulnerable to both me-

chanical damage and chemical attack than that of the GCE, it is

more easily regenerated: often it suffices merely to rub the tip of

the electrode backwards and forwards a few times against the

surface of a sheet of ordinary, smooth paper, or alternatively the

upper layer of carbon paste can be renewed (see below) and

then polished in this manner.

Filling

Carefully remove the old carbon paste from the approximately

2.5 mm deep cavity with a suitable sharp instrument. Take care

not to damage the platinum contact wire at the bottom of the cav-

ity. Using a small spatula, fill bit by bit with the fresh 6.2801.020

carbon paste and tamp well down using the 6.2826.000 stopper.

The cavity should be filled over the brim with paste. Now rub off

the excess paste by continuous circular motion against a sheet of

paper with a hard, smooth, non-porous surface; the circular mo-

tion plus the light pressure will polish the surface of the paste,

which must be smooth and shiny when the operation is finished.

There should be no cracks in the peripheral zone where the paste

is up against the plastic material of the electrode shaft.

Metal electrodes

Ordering designations 6.1204.120 Pt tip

6.1204.130 Ag tip

6.1204.140 Au tip

+ 6.2103.110 Contact pin

Chemical resistance

good resistance to acetonitrile, methanol, wa-

ter, etc.

Voltage range

Pt: (–1200)

0

+200 mV (pH < 7)

Ag: (–800)

0

+400 mV (pH > 7)

Au: (–400)

0

+1200 mV (pH < 7)

Regeneration

Sprinkle a little aluminium oxide powder (Al

2

O

3

, grain size 0.3

µ

m)

on the polishing cloth (ordering number of both these accesso-

ries: 6.2802.000) and moisten with distilled water. Hold the elec-

trode vertically and polish with a circular motion for about

20

60 s, exerting only light pressure. Rinse with distilled water

and clean with dry cloth. A metal electrode should always be pol-

ished for a short while immediately before being fitted to the de-

tector cell.