2 influence on detector signal – Metrohm 791 VA Detector User Manual

Page 22

background image

3 Operation

791 VA Detector

16

Structural formula

Classes/substances

Polarization voltage

R SH

Mercaptans

+

800

mV

Miscellaneous

Ascorbic acid
Carotenes
Purine derivatives
Vitamin A

+ 800

mV

+ 800

mV

+

1000

mV

+

1000

mV

R

O

O

Quinones

Vitamin K

1

– 700

mV

R NO

2

Nitro compounds

Nitrophenol
Nitroglycerin
Isosorbidedinitrate

– 700

mV

– 800

mV

– 800

mV

R

Inorganic anions

Nitrite
Sulfide
Thiosulfate
Chlorite
Bromide
Iodide
Cyanide
Thiosulfate
Thiocyanate

+

1100

mV

+ 0

mV

+

1100

mV

+

1100

mV

+ 150

mV

+ 150

mV

+ 0

mV

+

1100

mV

+

1200

mV

3.1.2

Influence on detector signal

The sensitivity of detection, i.e. the height of the detector signal and the

background current in the electrochemical detection depend on the fol-

lowing parameters:

polarization voltage

temperature (approx. 1.5 %/°C)

flow rate of eluent

surface of electrodes and interior of cell

Eluent composition

(pH value, type and concentration of conductive salt,

type and concentration of organic components, etc.)

Each change of one of these parameters will change the background

current and the detector signal. Therefore, every parameter change

needs a stabilization time to elapse; this is normally a few minutes for