2 influence on detector signal – Metrohm 791 VA Detector User Manual
Page 22

3 Operation
791 VA Detector
16
Structural formula
Classes/substances
Polarization voltage
R SH
Mercaptans
+
800
mV
Miscellaneous
Ascorbic acid
Carotenes
Purine derivatives
Vitamin A
+ 800
mV
+ 800
mV
+
1000
mV
+
1000
mV
R
O
O
Quinones
Vitamin K
1
– 700
mV
R NO
2
Nitro compounds
Nitrophenol
Nitroglycerin
Isosorbidedinitrate
– 700
mV
– 800
mV
– 800
mV
R
Inorganic anions
Nitrite
Sulfide
Thiosulfate
Chlorite
Bromide
Iodide
Cyanide
Thiosulfate
Thiocyanate
+
1100
mV
+ 0
mV
+
1100
mV
+
1100
mV
+ 150
mV
+ 150
mV
+ 0
mV
+
1100
mV
+
1200
mV
3.1.2
Influence on detector signal
The sensitivity of detection, i.e. the height of the detector signal and the
background current in the electrochemical detection depend on the fol-
lowing parameters:
•
polarization voltage
•
temperature (approx. 1.5 %/°C)
•
flow rate of eluent
•
surface of electrodes and interior of cell
•
Eluent composition
(pH value, type and concentration of conductive salt,
type and concentration of organic components, etc.)
Each change of one of these parameters will change the background
current and the detector signal. Therefore, every parameter change
needs a stabilization time to elapse; this is normally a few minutes for