background image

2 Installation

791 VA Detector

8

2.2 Mains

connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with a mains voltage set wrongly and/or

wrong mains fuse, there is a danger of fire!



2.2.1

Setting the mains voltage

Before switching on the 791 VA Detector for the first time, check that the

mains voltage set on the instrument (see Fig. 3) matches the local

mains voltage. If this is not the case, you must reset the mains voltage

on the instrument as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 26 of the

791 VA Detector.

2 Remove fuse holder

Using a screwdriver, loosen fuse holder 12 and take out com-

pletely.

3 Change mains voltage

Completely remove voltage selection insert 13 by hand, rotate it

through 180° and reinsert it. The required mains voltage (115 or

230 V) must now be visible from the front.

4 Check fuses

Carefully take both fuses out of fuse holder 12 and check their

specifications:

100

240 V 1 A (slow-blow) Metrohm-No.

U.600.0016

5 Insert fuses

Change both fuses if necessary and reinsert in fuse holder 12.

6 Install fuse holder

Push fuse holder 12 back into the opening of 791 VA Detector

by hand until it clicks into place properly.