2.4 6.5303.030 ELCD cell for IC

791 VA Detector

11

2.4

6.5303.030 ELCD cell for IC

2.4.1

Assembling the ELCD cell

The reference electrode and working electrode must be inserted in the

6.5303.030 ELCD Cell supplied with the 2.791.0020 VA Detector. The

cell is then screwed into the detector block built into the 733 IC Separa-

tion Center. Proceed as follows (see Fig. 5 and Fig. 6):

1 Insert reference electrode

Reference electrode 39 (6.0727.000, details see section 3.2.2)

included in the supply is already filled with c(KCl) = 3 mol/L and

protected with a cap. It is assembled as follows (see Fig. 5):

Screw off the cap of electrolyte storage vessel 41 of reference

electrode 39. Normally PTFE gasket 31 remains in the cap

and can be left there.

Insert one of the supplied PTFE gaskets 31 into the corre-

sponding opening of cell body 32 and screw reference elec-

trode 39 into this opening.

2 Insert working electrode

Working electrode 28 is not supplied with the 791 VA Detector

and must be ordered separately. Details about the working

electrodes and information about pretreatment (e.g. polishing)

can be found in section 3.2.1. A working electrode is assembled

as follows (see Fig. 5):

Insert working electrode 28 in screw nipple 29.

Push one of the supplied PTFE gaskets 31 sufficiently far

along the working electrode from below.

Carefully insert working electrode 28 into the corresponding

opening of cell body 32 from above and press it in until it

clearly clicks into position.

Push screw nipple 29 downwards and screw it in.

3 Connect ELCD cell to the detector block

The ELCD cell is screwed onto the 1.733.0X10 Detector block

mounted inside the 733 Separation Center. Proceed as follows

(see Fig. 5 and Fig. 6):

Screw cell body 32 onto detector block 42 with the help of

the two red knurled screws 43.

Push one of the supplied PTFE gaskets 31 over screw nipple

33 and screw this into the corresponding opening of cell

body 32.

Place a screw nipple 36 and a PTFE gasket 35 on the outlet

end of the outlet capillary 37 permanently connected to de-

tector block 42.

Insert outlet capillary 37 with screw nipple 36 in place into

screw nipple 33 and press in outlet capillary 37 until it

reaches the stop.