750 Autosampler
49
5
Appendix
5.1
Technical data
Sampling
Method
Positive displacement principle
Capacity
128 sample vials of 730
µ
L each
Number of injections per vial max. 4 with sample loops
≤
10
µ
L
max. 3 with sample loops >10
µ
L
Max. loop size
50
µ
L with 3 injections per sample
100
µ
L with 2 injections per sample
300
µ
L with 1 injection per sample
Carryover
< 0.25 % without rinsing
< 0.004 % with rinsing
Control elements
Display
LCD, 4 lines, each of 20 characters
(character height 4 mm)
Keypad
Membrane keypad made of polyester
with 20 keys
Mains switch
At rear of instrument
Mains connection
Voltage
100/120 V
±
10 %
230/240 V
±
10 %
autoselectable
Frequency
50/60 Hz
Power consumption
80 VA
Fuses
100/120 V: 2 A (rapid-action, with low rupturing
capacity, 3AG)
230/240 V: 1.25 A (rapid-action, with low
rupturing capacity, 5
×
20 mm)
Interfaces
Contact closure strips
Outputs for control of external devices,
inputs for remote control (see section 5.2.1)
RS232 interface
Data communication in accordance with RS232C
(see section 5.2.2)