5.1 Technical data

750 Autosampler

49

5

Appendix

5.1

Technical data

Sampling

Method

Positive displacement principle

Capacity

128 sample vials of 730

µ

L each

Number of injections per vial max. 4 with sample loops

10

µ

L

max. 3 with sample loops >10

µ

L

Max. loop size

50

µ

L with 3 injections per sample

100

µ

L with 2 injections per sample

300

µ

L with 1 injection per sample

Carryover

< 0.25 % without rinsing

< 0.004 % with rinsing

Control elements

Display

LCD, 4 lines, each of 20 characters

(character height 4 mm)

Keypad

Membrane keypad made of polyester

with 20 keys

Mains switch

At rear of instrument

Mains connection

Voltage

100/120 V

±

10 %

230/240 V

±

10 %

autoselectable

Frequency

50/60 Hz

Power consumption

80 VA

Fuses

100/120 V: 2 A (rapid-action, with low rupturing

capacity, 3AG)

230/240 V: 1.25 A (rapid-action, with low

rupturing capacity, 5

×

20 mm)

Interfaces

Contact closure strips

Outputs for control of external devices,

inputs for remote control (see section 5.2.1)

RS232 interface

Data communication in accordance with RS232C

(see section 5.2.2)