3 Operation
750 Autosampler
34
3.6
Method procedure
RUN/STOP
A previously programmed method can be started by pressing
the
by selection of the menu item "
Run Method
" in the method
menu. In both cases the following display appears:
Enter Method Number
# N
1
…
10
Select method
Entry of the method number N which is to be started.
After confirmation with
method is started. While the program is running the
following display appears:
METHOD STEP METHTIME
NN MM XXX.X
VIAL INJ RTIME
nnn m xxx.x
Status display for running method
While the program is running the following data are
displayed and continually updated:
METHOD
Number NN of current method
STEP
Number MM of program step
METHTIME
Total elapsed time since method start
VIAL
Number nnn of current vial
INJ
Number m of current injection
RTIME
Run time of current sample
When the last program step has been completed the
transfer needle is returned to the park position. The
following display appears:
METHOD
COMPLETED
Method completed
The selected method has been correctly completed.
The instrument is ready for a method to be restarted or
to return to the basic condition.
METHOD STOPPED
Method stopped
A running program can be stopped at any time with
the
its position and the display shown alongside appears.
The three following possibilities now exist:
Continue method run.
Abort method.
Edit the current method, then re-
turn to this selection.