4 Notes – Maintenance – Faults
750 Autosampler
42
Fig. 11:
Transfer needle replacement
4.2.3
Sample preparation
Experience has shown that most problems are caused by solids in the
sample solutions. Such particles can block the needle, transfer tubing
and the sample loop and also permanently damage the surfaces of the
injection valve seals. This is the reason why all samples must be filtered
before they are filled into the sample vials (procedure see chap. 4.1.2).
Solids may also be formed by the precipitation of buffer salts in the IC
system. With buffered samples it is recommended that the sample loop
feed is rinsed with a large volume of pure water after sample injection.
For example, this can be achieved by filling several (e.g. 5) sample vials
in sequence with distilled water after each sample and programming
them as rinsing samples without run time. This procedure also avoids
corrosion damage to the injection needle and the crystallization of salts.
17
17 Needle holder
55 Transfer needle
16
16 Ferrule
PEEK compression fitting 18
18
Nut 19
19
Connection screw 21
21
Centering sleeve 24
24
Clamping ring 22
22
Holder 23
23
Holder 20
20
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (163 pages)
- 756 KF Coulometer (162 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)